Gwyscope documentation

open hardware SPM controller with advanced sampling support

User Tools

Site Tools


mfm

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
mfm [2025/05/12 16:42] pklapetekmfm [2025/05/12 16:45] (current) pklapetek
Line 3: Line 3:
 Magnetic force microscopy (MFM) is a dual-pass technique in which a magnetically coated probe is used to get information about the magnetic stray field distribution above the surface. Magnetic force microscopy (MFM) is a dual-pass technique in which a magnetically coated probe is used to get information about the magnetic stray field distribution above the surface.
  
-MFM measurements are performed in tapping mode, so to perform MFM with Gwyscope, first set up the [[tapping|tapping mode]]. When done, set the [[lift|dual path measurement]] and let the microscope collect amplitude and phase in some distance above surface, typically in tens of nanometers.+MFM measurements are performed in tapping mode, so to perform MFM with Gwyscope, first set up the [[tapping|tapping mode]]. When done, set the [[lift|dual path measurement]] and let the microscope collect amplitude and phase in some distance above surface, typically in tens of nanometers. When number of lifts is non-zero, additional channels can be selected for the previews (both for image and graph) and the data measured in lift can be visualized.
  
-An example of Gwyscope running a MFM measurement on a MFM calibration sample (prepared by IFW Dresden) is shown below.+An example of Gwyscope running a MFM measurement on a MFM calibration sample (manufactured by IFW Dresden) is shown below.
  
 +{{ :mfm_scanning.png?direct&600 |}}
mfm.1747060968.txt.gz · Last modified: 2025/05/12 16:42 by pklapetek