| Line 1: | Line 1: | ||
| ====== Sideband KPFM ====== | ====== Sideband KPFM ====== | ||
| + | |||
| + | Sideband KPFM is in principle similar to [[fmkpfm|FM KPFM]], again using combination of mechanical and electrical oscillation, | ||
| + | |||
| + | To set up the hardware for FM-KPFM operation with Gwyscope, use the same settings as for [[amkpfm|AM KPFM]]. | ||
| + | |||
| + | To set up the operation, first measure the second harmonic resonance frequency using Spectral probe response dialogue. For rectangular cantilevers, | ||
| + | Then, pick " | ||
| + | |||
| + | Then the KPFM dialogue (see [[amkpfm|AM KPFM]] description for more details) can be used to select the right signal for feedback by adjusting the phase or altering the excitation frequency. | ||
| + | |||
| + | |||
sideband.1740476691.txt.gz · Last modified: 2025/02/25 10:44 by pklapetek